A new RP-HPLC approach for estimation of potential impurities of Fosamprenavir - method development and validation
BMC Pharmacology and Toxicology volume 26, Article number: 60 (2025)
Abstract
The current work aims to develop a reliable and robust RP-HPLC method for analyzing Fosamprenavir and its potential impurities, including isomer, amino, propyl, nitro, and Amprenavir. The method used a Zobrax C18 column with a mobile phase of 0.1% V/V orthophosphoric acid in water and acetonitrile in gradient elution at a flow rate of 1 mL/min to accomplish efficient separation with detection at 264 nm and column temperature of 30 ± 20C. A diluent with a 1:1 water-to-acetonitrile ratio was used to prepare standard and sample solutions. The developed approach was validated as per ICH Q2(R1) guidelines. Fosamprenavir, Amino, Propyl, Isomer, Nitro impurities, and Amprenavir impurities were eluted at retention time (RT) of 5.3 min, 2.3 min, 4.3 min, 4.7 min, 8.1 min and 8.6 min correspondingly with good resolution within a 10-minute run time. Method validation confirmed system suitability, linearity (R² = 0.999), good sensitivity (LOD/LOQ), specificity, precision (% RSD: 0.5–1.7), accuracy (% recovery: 90.9-104.3%), and robustness. The optimized approach excelled existing methods in lower retention time, run time, sensitivity, and linearity for all potential impurities, making it a novel and trustworthy method for monitoring Fosamprenavir drug quality and assessing stated impurities. The established method allows precise measurement of Fosamprenavir and related substances, supporting drug safety and regulatory compliance.
Graphical Abstract

Introduction
Fosamprenavir, a derivative of amprenavir, is categorized within the protease inhibitor class of pharmaceuticals, pivotal in highly active antiretroviral therapy (HAART) regimens for HIV/AIDS management [1,2,3]. Its inception in 2003 stemmed from the imperative to surmount amprenavir’s bioavailability and dosing intricacies [2]. Fosamprenavir operates akin to amprenavir by impeding the functionality of the HIV protease enzyme, pivotal for viral replication [2,3,4]. Through this inhibition, Fosamprenavir disrupts the processing of viral polyproteins, impeding the formation of mature, infectious HIV particles [2,3,4]. Consequently, viral replication is curtailed, mitigating disease advancement by sustaining low viral levels [4,5,6]. Fosamprenavir is sparingly soluble in water but freely soluble in Acetonitrile, DMF, DMSO, ethanol, and methanol [7, 8]. Chemically, Fosamprenavir is [(3 S)-oxolan-3-yl] N-[(2 S,3R)-4-[(4-aminophenyl)sulfonyl-(2-methylpropyl)amino]-1-phenyl-3-phosphonooxybutan-2-yl]carbamate with molecular formula of C25H36N3O9PS [7, 8].
Impurity profiling is the process of identifying and quantifying impurities in a pharmaceutical substance or product. Impurities can come from various reasons, including the manufacturing process, storage conditions, and the drug’s degradation over time [9]. The presence of impurities can affect a drug’s safety and efficacy. Impurity profiling can be performed using a range of analytical techniques, such as high-performance liquid chromatography (HPLC), mass spectrometry (MS), and nuclear magnetic resonance (NMR) [9,10,11]. HPLC is a potent method for separating, detecting, and quantifying impurities in pharmaceutical compounds. Its adaptability and capacity to combine with other analytical methods make it an essential tool in impurity profiling to ensure drug safety and quality [10, 11].
The potential impurities of Fosamprenavir during synthesis or storage include isomer, amino, propyl, nitro, and Amprenavir impurities [12, 13]. An in-depth exploration of the literature regarding analytical methods for Fosamprenavir ensures that few UV and HPLC methods are used to estimate Fosamprenavir in drug substances and products [14,15,16,17,18,19]. Along with the HPLC method, few LC-MS methods are reported for the estimation of Fosamprenavir in biological samples and the characterization of metabolites and forced degradants [20, 21]. Only one HPLC method was reported for concurrently identifying and estimating Fosamprenavir and its isomer, amino, and Amprenavir impurities [13]. Along with the prior method, one more liquid chromatographic method is reported for the estimation of potential impurities of Fosamprenavir [22]. Propyl and Nitro impurities of Fosamprenavir are also potential impurities stated in the previous method. A competent, accurate, specific, and sensitive HPLC method is essential for identifying and estimating all potential process impurities in trace levels. Hence, an attempt was made to make a good and proficient RP-HPLC method for simultaneous analysis of Fosamprenavir and its process impurities (Amino, Propyl, Isomer, Nitro impurities, and Amprenavir). The chemical structures of Fosamprenavir and stated impurities are shown in Fig. 1.
Materials and methods
The pure forms of Fosamprenavir (99.98%), Amino, Propyl, Isomer, Nitro impurities, and Amprenavir impurities were obtained as gift samples from Icon Laboratories, Andhra Pradesh, India. The potencies of all the impurities ranged from 99.87 to 99.93%. The HPLC-grade methanol and acetonitrile were purchased from a local distributor of Finar Chemical Limited, India. The executive of the present method development and validation, WATERS HPLC with PDA detector integrated with Empower-2 software, was used.
Method development
The current method was developed by using WATERS HPLC equipped with a binary pump and PDA detector. Initially, method development started with an isocratic mobile phase of various ratios of 0.1% orthophosphoric acid (OPA) and acetonitrile in 50:50, 60:40, and 70:30. During those conditions, good resolution between Fosamprenavir, Isomer and Propylimpurites were too low. Amino and Nitro impurities were not eluted with the isocratic elution. Finally, the Zobrax C18 column (100 × 4.6 mm, 5 μm) was used in conjunction with a mobile phase of 0.1% v/v OPA in water and acetonitrile in gradient elution mode opted to attain good resolution among analytes (Fosamprenavir and impurities) (Table 1). The mobile phase was fed to the column at a flow rate of 1 ml/min, and eluted chemicals were detected at a wavelength of 264 nm. The sample input and column temperature were kept at 30 ± 20C. A diluent with a 1:1 water-to-acetonitrile ratio was used to prepare standard and sample solutions.
Preparation of standard solution
The standard solution was prepared by transferring 1 mg of each Fosamprenavir, Amino, Propyl, Isomer, Nitro impurities, and Amprenavir impurity into 100 ml of a volumetric flask. 50 ml of diluent (1:1 ratio of acetonitrile and water) was added and sonicated for 5 min; the remaining volume was adjusted with the same diluent to get a solution of 10 ppm. 0.1 ml of the above-stated solution was further diluted with the same diluent to attain a solution of 1ppm.
Preparation of sample solution
Tablet powder equivalent to 1 mg of each Fosamprenavir 100 ml of volumetric flask. 50 ml of diluent (1:1 ratio of acetonitrile and water) was added and sonicated for 5 min; the remaining volume was adjusted with the same diluent to get a solution of 10 ppm. 0.1 ml of the above-stated solution was further diluted with the same diluent to attain a solution of 1ppm.
Method validation
The present method was validated as per ICH Q2(R1) provisions [23,24,25,26].
System suitability
The system suitability of the current RP- HPLC method was confirmed by injecting six subsequent injections of a standard solution consisting of 1 ppm of Fosamprenavir, Amino, Propyl, Isomer, Nitro impurities, and Amprenavir impurity each injected into the HPLC system. The chromatograms were interpreted to assess the theoretical plate count, USP resolution, tailing factor, and %RSD of the peak areas for obtained peaks.
Linearity
The linearity for Fosamprenavir and its stated relative impurities was performed for a concentration series of 25–150% level of standard solution (0.5,0.75,1,1.25 and 1.5 ppm). Each concentration level was injected three times (n = 3), and the mean peak area was considered to plot a linear graph between concentration and peak area to compute the R2 value.
Sensitivity
The signal-to-noise (S/N) ratio method is commonly used to determine the LOD (Limit of detection) and LOQ (Limit of Quantification). The LOD represents the concentration or amount of analyte yielding a signal three times the standard deviation of the background noise (S/N ratio of 3:1), and the LOQ represents the concentration or amount of analyte producing a signal ten times the standard deviation of the background noise. This method entails detecting the background noise level without an analyte and quantifying the signal produced by a known low analyte concentration. The LOD is then computed as a signal’s concentration three times the noise’s standard deviation. In contrast, the LOQ is calculated similarly for a signal ten times the standard deviation of the noise. The reliability of LOD and LOQ of stated impurities and Fosamprenavir were confirmed by analysing LOD and LOQ concentrations for 5 repetitive injections to assess the %RSD.
Specificity
Specificity in an HPLC (High-Performance Liquid Chromatography) method refers to its ability to accurately determine the analyte desired in the presence of potentially interfering chemical substances. It ensures that the process can identify and measure the target chemical substances without interference from other substances in the sample matrix. In the current method, specificity was performed by injecting successive injections of individual impurity solutions, a solution of all impurities spiked with Fosamprenavir solution and Fosamprenavir sample solution into the HPLC system. The recorded chromatograms were interpreted to identify interferences among the RT of all impurities, interferences of RT of Fosamprenavir with RT of individual impurities, and interferences with placebo towards Fosamprenavir and impurities.
Precision
The optimized method’s system precision was validated by injecting a standard solution consisting of Fosamprenavir, Amino, Propyl, Isomer, Nitro impurities, and Amprenavir impurity in 6 repeated injections. The %RSD was statistically computed for the results in peak areas of Fosamprenavir and stated impurities in the replicated injections. The method precision of the optimized method was done by injecting Fosamprenavir sample solution spiked with 0.1% of each impurity (Amino, Propyl, Isomer, Nitro, and Amprenavir impurity). The % RSD for the %recovery of each spiked solution was calculated for six consecutive injections.
Accuracy
To ensure the accuracy of the current HPLC method, a % recovery procedure was chosen, in which Fosamprenavir sample solutions were spiked with all the related substances or impurities at different concentration levels of LOQ, 50, 100, and 150% of each. Three serial injections of each spiked solution were introduced into HPLC, and the % mean recovery of each impurity in the spiked solution was.
Robustness
To confirm the robustness of the HPLC method, slight variations were made in method conditions with intention. Small, deliberate changes were made to parameters including flow rate (± 0.1 ml/min), column temperature (± 2 and maximum wavelength (± 2 nm). The %RSD of the obtained peaks was determined.
Results
Optimized method
After several trials, a method with Zobrax C18 column (100 × 4.6 mm, 5 μm), a mobile phase of 0.1% v/v OPA in water and acetonitrile in gradient elution mode, the flow rate of 1 ml/min, and detection wavelength of 264 nm is used. These chromatographic conditions can be used to separate Fosamprenavir, Amino, Propyl, Isomer, Nitro impurities, and Amprenavir impurities at retention time (RT) of 5.3 min, 2.3 min, 4.3 min, 4.7 min, 8.1 min and 8.6 min correspondingly with good resolution (Fig. 2).
Method validation
The system suitability of the analytical method was ensured by assessing the % RSD, USP tailing, USP plate count, and resolution. All the parameters’ values were aligned with Q2 specifications ICH guidelines (Table 2). The R2 value for the Fosamprenavir and stated impurities were assessed to be 0.999, corresponding to the stated concentration series (Table 3; Fig. 3). This demonstrates the effectiveness of the suggested methodology in exhibiting notable linearity throughout the designated concentration series. The LOD and LOQ results ascertained by the S/N ratio method for all analytes were in the very low range, demonstrating the remarkable sensitivity of the stated method for Fosamprenavir and its impurities. The sensitivity results, including LOD and LOQ with their S/N ratio and %RSD of peak responses, were stated in Table 4. The chromatogram representing LOQ levels was shown in Fig. 4. There was no interference at the RT Fosamprenavir with blank and stated impurities in the recorded chromatograms (Fig. 5), which disclose the specificity of the current HPLC approach towards the Fosamprenavir and stated impurities of Fosamprenavir. The %RSD results of system precision (peak responses of all analytes in standard solution) and method precision (% recovery of impurities from spiked solutions) were in the range of 0.5 to 1.7, confirming the precision of the developed approach in accordance with ICH regulations (Table 5). The % mean recovery of each impurity in spiked solutions of Fosamprenavir was computed to be in the range of 90.9–104.3% (Table 6), which powerfully reveals the accuracy of the stated method. The method can maintain consistent results even after making slight intentional alterations to the method’s conditions. The robustness of the approach is demonstrated by the %RSD values of peak responses, which were within the permissible limits established by the ICH (Table 7).
Discussion
Related substances (RS) HPLC procedures have become essential in pharmaceuticals for detecting and quantifying impurities and assuring regulatory compliance, product safety, stability, and uniformity. The acceptable limits for impurities in Active Pharmaceutical Ingredients (API) as specified by various regulatory agencies are less than 0.1% for each impurity and less than 1% for total impurities [25]. They optimize processes, help evaluate risks, and promote research and development, ensuring the quality and efficacy of pharmaceuticals. The reported potential impurities of Fosamprenavir include isomer, amino, propyl, nitro, and Amprenavir impurities. Only two RS-HPLC methods have been reported for detecting and quantifying Fosamprenavir impurities. Only isomer, amino, and Amprenavir impurities of Fosamprenavir were identified and quantified using one method. In the other method, impurities 2 and 5 of Fosamprenavir were estimated. The present method is superior to the reported methods in terms of sensitivity and linearity range. The LOD and LOQ values of impurities and Fosamprenavir in the current method are higher than in the earlier methods [13, 22]. In the reported method, LOD and LOQ values were found to be isomer (0.06, 0.17 ppm), amino (0.07,0.2 ppm), and Amprenavir (0.1, 0.3 ppm) impurities [13]. In the reported method, all potential impurities were not estimated concurrently. The current method can estimate all the stated impurities simultaneously with a run time of 10 min and RT of 5.3 min, 2.3 min, 4.3 min, 4.7 min, 8.1 min, and 8.6 min for Fosamprenavir, Amino, Propyl, Isomer, Nitro impurities, and Amprenavir impurity correspondingly with good resolution. The validation parameter results satisfied the ICH Q2 acceptance criteria, ensuring the method’s competency.
Conclusion
An easy and efficient RP-HPLC method estimation of Fosamprenavir and its isomer, amino, propyl, nitro, and Amprenavir impurities in bulk and tablet form was developed. Using optimized method conditions, shorter elution time, reproducible precision, and high sensitivity were achieved. The method consists of reproducible specificity to assess Fosamprenavir and its impurities. Hence, the developed method has reasonable application in the pharmaceutical industry.
Data availability
All data provided within the manuscript.
Abbreviations
- OPA:
-
Orthophosphoric acid
- LOD:
-
Limit of detection
- LOQ:
-
Limit of quantification
- RSD:
-
Relative standard deviation
- SD:
-
Standard deviation
References
Chapman TM, Plosker GL, Perry CM. Fosamprenavir: a review of its use in the management of antiretroviral therapy-naive patients with HIV infection. Drugs. 2004;64:2101–24.
Crowe SM, Katlama C, Murphy R. Fosamprenavir. Kucers’ The Use of Antibiotics. CRC; 2017. pp. 4104–20.
Torres HA, Arduino RC. Fosamprenavir calcium plus Ritonavir for HIV infection. Expert Rev Anti-infect Ther. 2007;5:349–63.
Wire MB, Shelton MJ, Studenberg S. Fosamprenavir: clinical pharmacokinetics and drug interactions of the amprenavir prodrug. Clin Pharmacokinet. 2006;45(2):137–68.
Arvieux C. Amprenavir or Fosamprenavir plus Ritonavir in HIV infection: pharmacology, efficacy and tolerability profile. Drugs. 2005;65:633–59.
Gatti F, Nasta P, Loregian A, Puoti M, Matti A, Pagni S, Carosi G. Unboosted Fosamprenavir is associated with low drug exposure in HIV-infected patients with mild–moderate liver impairment resulting from HCV-related cirrhosis. J Antimicrob Chemother. 2009;63:575–8.
National Center for Biotechnology Information. (2024) PubChem compound summary for CID 131536. Fosamprenavir. Retrieved May 21, 2024.
Cordeiro CF, Bettio I, Trevisan MG. Studies on the characterization and polymorphic stability of Fosamprenavir. Acad Bras Cienc. 2020;92:e20181021.
Shah SR, Patel MA, Naik MV, Pradhan PK, Upadhyay UM. Recent approches of impurity profiling in pharmaceutical analysis: A review. Int J Pharm Sci Res. 2012;3(10):3603.
Gorog S. The importance and the challenges of impurity profiling in modern pharmaceutical analysis. TrA C. 2006;25:755–7.
Ingale SJ, Sahu CM, Paliwal RT, Vaidya S, Singhai AK. (2011) Advance approaches for the impurity profiling of pharmaceutical drugs: Int J Pharm Sci Res 2(7).
Arora SK, Shabade SS, Kumar G, Ray PC, Singh GP. U S Patent No. 2014;8:877947.
Nekkala K, Kumar JS, Ramachandran D. Stability indicating RP-HPLC method for quantification of impurities in Fosamprenavir calcium drug substance. J Pharm Sci. 2019;11:712–7.
Pekamwar SS, Bhavar GB, Aher KB, Kakad SJ. Validated UV spectrophotometric and HPTLC method for determination of Fosamprenavir calcium in pharmaceutical formulation. Austin J Anal Pharm Chem. 2015;2:1049–55.
Rossi RC, Moura JGL, Mossmann V, Weimer P, Fröehlich PE. A new, rapid and simple RP-HPLC method for stability quantification of a protease inhibitor in tablets. J Chromatogr Sci. 2022;60:671–7.
Chilukuri M, Narayanareddy P, Hussianreddy K. Stability-indicating HPLC method for determination of Fosamprenavir calcium. J Chromatogr Sci. 2014;52:781–7.
Rao NM, Gowrisankar D. Development and validation of stability indicating RP-HPLC method for Estimation of Fosamprenavir calcium in pure and pharmaceutical dosage forms. Asian J Chem. 2015;27:3484.
Anitha D, Manoranjani M, Parameswararao K. (2021) Affordable reverse phase fluid chromatographic strategy for the assessment of Fosamprenavir in tablets. Turk J Qual Inq 12(6).
Kurbanoglu S, Ashrafi A, Vytras K, Uslu B, A Ozkan S. Determination and validation of the antiviral drugs using reverse phase high performance liquid chromatographic method. Curr Drug Ther. 2014;9:256–60.
Thimmaraju MK, Mondal P, Venu K, Babu GS, Rajkumar G, Padmaja B. Quantification of Fosamprenavir in spiked human plasma using liquid chromatography–electrospray ionization–tandem mass spectrophotometry–application to Pharmacokinetic study. Egypt Pharm J. 2019;18:149–59.
Susmitha K, Menaka M. LC-ESI-MS/MS method for the Fosamprenavir quantification bioanalytical method development and validation for the quantification Fosamprenavir in human plasma by LC-ESI-MS/MS. Indian J Pharm Educ Res. 2022;56:S599–604.
Singh GVS, Reddy DRS, Divakar TE. (2019) Stable liquid chromatography method for the resolution and quantification of related impurities of Fosamprenavir in bulk and pharmaceutical formulations. Indian Drugs 56(9).
Ettaboina SK, Katakam LN, Dongala T. Development and validation of a stability-indicating RP-HPLC method for the determination of erythromycin-related impurities in topical dosage form. Pharm Chem J. 2022;56:131–7.
Ettaboina SK, Nakkala K, Chathalingath N. (2022) Development and validation of rapid RP-HPLC method for determination of deferasirox in bulk and tablet dosage forms. Asian Journal of Pharmaceutical Analysis. 2022: 73– 7.
ICH, Quality Guidelines ICH. accessed January 18,. https://www.ich.org/page/quality-guidelines (2024).
Ettaboina SK, Nakkala K. (2021). Development and validation of new analytical methods for the simultaneous estimation of levamisole and Albendazole in bulk and tablet dosage form by RP-HPLC. Journal of Pharmaceutical Analytics and Insights. 2021:1–6.
Acknowledgements
The authors thank the Department of Pharmaceutical Analysis, GITAM Deemed to be University, Hyderabad, Malla Reddy Institute of Pharmaceutical Science and Saveetha medical college and Hospital, Saveetha institute of medical and technical sciences for continuous support.
Funding
The authors declare that no funds, grants, or other support were received during the preparation of this manuscript.
Author information
Authors and Affiliations
Contributions
All authors contributed to the study’s conception and design. RG, MVN, PKJ, VM, KP, SP performed material preparation, data collection, and analysis. RG, PPR, GSB wrote the first draft of the manuscript, VKN, KKK supervised the entire work and edited the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Godela, R., Nelson, V.K., Nuli, M.V. et al. A new RP-HPLC approach for estimation of potential impurities of Fosamprenavir - method development and validation. BMC Pharmacol Toxicol 26, 60 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s40360-025-00892-5
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s40360-025-00892-5